loading

 Rika Sensor is a weather sensor manufacturer and environmental monitoring solution provider with 10+ years of industry experience.

How to Calibrate a pH Sensor?

Did you know that pH sensors work on the fundamental law of electrochemistry? A perfect pH sensor will produce 59.16 mV for every 1.0 pH unit change in solution. However, with time, this voltage can change owing to the buildup of dirt or impurities on the electrode. The result is inaccurate readings and drift in values. Therefore, we need to calibrate the pH sensor at regular intervals to ensure that our results remain reliable.

 

Calibrating a pH sensor requires an understanding of its working mechanism. There are different types of pH sensors, which can have varying responses to changes in pH. We must understand the differences and apply the method of calibration with attention to ensure precise results. High-end sensors like the RK500-12 series from Rika have features that help maintain consistent results through a low-impedance glass membrane.

 

Tons of features can make water pH sensors easier to calibrate and maintain. In this comprehensive guide, we will start with the basics on what a pH sensor is, why you need to calibrate a pH sensor, and when you need to calibrate a pH sensor. Provide the latest information on pH sensors, focusing on the guideline's highlights, specifically the step-by-step process for calibrating a pH sensor to achieve accurate results. Let's begin learning!

How to Calibrate a pH Sensor? 1

What is a pH Sensor?

Definition

pH sensors measure the level of acidity or alkalinity of the liquid. A scale that runs from 0 to 14 is used for measuring pH. Here, 14 means an alkaline solution, while 0 means an acidic solution. A neutral water solution will have 7 pH of 7. A pH sensor simply checks the solution and measures its pH level. pH is the potential/power of hydrogen.

 

Working Principle of pH Sensor

The working mechanism of a pH sensor is based on electrochemical measurement. It is vital to understand the working principle so that the relevant method can be chosen for calibrating the pH sensor. To detect pH, a sensor needs a glass electrode, a reference electrode, and a signal converter to produce results. Here are the roles of each component:

  • Glass Electrode: It is the bulb-shaped glass membrane typically located at the tip of a pH sensor. The purpose is to come in contact with the sample water. The glass is made with low-impedance material to ensure higher sensitivity. When it comes into contact with the water, it creates a potential difference proportional to the hydrogen ions (H⁺) in the solution. The Nernst equation drives the whole phenomenon:

E = E0 + (2.303 . RT/nF) . log10 [H⁺]

E is the measured potential, E0 is the standard potential, R is the gas constant, T is the temperature, n is the charge, and F is Faraday’s constant.

 

  • Reference Electrode: The reference electrode provides a stable potential against which the glass electrode potential is compared. It is usually a wire dipped into a KCI solution. The KCI solution comes in contact with the sample water through a membrane.
  • Signal Converter: pH is the potential difference between the glass and reference electrode. The potential difference is recorded against standard solutions with known pH. Finally, a signal in the form of potential is converted into RS-485 and 4-20mA.

Types of pH Sensors

Glass Electrode pH Sensors

The reference electrode is not present inside the probe. It can be inserted separately into the sample water solution. It only has the glass electrode that produces a potential against hydrogen ions (H⁺).

How to Calibrate a pH Sensor? 2 

Example: Rika’s RK500-12 series Type-A1 for conventional water and Type-B3 for high-temperature environments

 

Combination Electrode pH Sensors

These sensors incorporate both the glass electrode and the reference electrode into the compact probe. They use the tube-within-a-tube concept. The inner tube has the glass bulb, and the outer tube has been filled with KCI solution for reference.

How to Calibrate a pH Sensor? 3 

Example: Rika’s RK500-12 series Type-C1 water pH level sensor

 

ISFET (Ion-Sensitive Field-Effect Transistor) pH Sensors

These utilize solid-state sensors that are sensitive to the H+ ion. It measures the pH change through its conductivity. These are resistant to breakage and suitable for high-pressure environments.

How to Calibrate a pH Sensor? 4 

 

Metal Oxide pH Sensors

Instead of using glass, metal oxide pH sensors use metal such as iridium oxide for detecting H+ and creating a potential. They offer robustness in harsh conditions and detect pH through redox reactions.

 

Submersible pH Sensors

These pH sensors have the capability to work while submerging the whole body is submerged in water. It is typically achieved using efficient sealing systems. These sensors will mention their IP rating using sealed connectors (M8/M16) and robust materials (Glass+ABS).

How to Calibrate a pH Sensor? 5 

Example: Rika’s RK500-12 series Type-B2 Submersible pH Sensor

 

Step-by-Step Guide to Calibrating a pH Sensor

Key Principles Before Starting

  • Frequency: Calibrate every 2-3 hours for lab use or based on inspection (1-2 days initially, then adjust per fouling/drift). For industrial-scale pH sensors, Rika recommends 3-6 months' routine checks.
  • Buffers: Use traceable, certified solutions (e.g., pH 4, 7, 10) spanning your measurement range, preferably ≥2 pH units apart. Discard the solution used for calibration after use. Ensure that the temperature matches the process (25°C standard).
  • Tools Needed: pH meter, buffers, distilled/deionized water, thermometer/RTD (for compensation), soft brush/HCl/NaOH for cleaning, tissue (no rubbing to avoid static).
  • Misconceptions: Expect nonlinearity (3 sections: 0-2, 6-8, 12-14 pH due to isoelectric point shifts). Activity coefficients are often unnecessary. Use standard addition for complex samples.
  • Safety: Handle acids/caustics with gloves. Avoid cyanide with HCl (toxic gas risk).

How to Calibrate a pH Sensor? 6 

Basic structure of a glass pH electrode, showing the glass membrane, internal reference, and electrolyte.

 

Step 1: Preparation and Initial Inspection

  • Gather Materials: Ensure buffers are fresh (e.g., pH 4.01, 7.00, 9.21/10.01. Use distilled water for rinsing. Have a thermometer if no built-in compensation.

Note: Rika's feature auto-adjustment of thermal resistance for 0-100°C in some models.

  • Inspect the Sensor: Remove from process/storage. Visually check for fouling, cracks, or low electrolyte (glass electrodes). For sensors like Rika's RK500-12, ensure the IP68 seal is intact.
  • Power On and Stabilize: Turn on the meter. Allow 1-3 minutes for the sensor output to hit equilibrium. For microprocessor-based meters (e.g., Rika's RS485 output), enter calibration mode.

Step 2: Cleaning the Sensor

Cleaning is mandatory before calibration to remove coatings or dirt that cause errors. Skip only if a quick check confirms no debris.

  • Rinse: Use warm distilled/tap water or a jet/spray to remove loose debris.
  • General Cleaning: Soak in detergent-water for 5 minutes. Gently scrub the bulb/reference with a soft brush. Ensure that the brush isn't abrasive.
  • For Specific Contaminants:
    • Alkaline/Scale: 5-10% HCl or vinegar soak (<5 min)
    • Acidic: Weak NaOH (<4%) soak
    • Oil/Organic: Detergent or compatible solvent
    • Inorganic/Organic: 0.1 mol/L HCl/NaOH for inorganics. Alcohol/acetone for organics, then rinse.
  • Rinse and Stabilize: Rinse with water; soak in 7 pH buffer or tap water for a few minutes to neutralize/stabilize. Dry by dabbing (no rubbing; Hamilton).
  • Avoid for Other Types: ISFET/metal oxide may not need electrolyte checks, but avoid abrasives.

Note: If readings remain off after the cleaning, proceed to calibration. If the electrode is heavily fouled, replace it.

 

Step 3: Perform a Calibration Check (Optional Quick Verification)

Before full calibration, verify if needed.

  • Immerse in Buffers: Place in pH 7 (offset) and pH 4/10 (span); note readings without adjusting.
  • Check Tolerance: If within ±0.1 pH, no calibration is needed—reinstall. (Rika offers ±0.01 pH accuracy).
  • Stirring: Stir gently for uniformity, then measure unstirred to avoid disrupting the Diffusion Layer (DL)/Transition Layer(TL).

Note: If out of tolerance, clean again or calibrate.

 

Step 4: Full Calibration (Two-Point Standard for Glass Electrodes)

For most applications, using a two-point standard is efficient and restores the accuracy of pH sensors. Microprocessor-based meters will auto-calculate slope/offset. You may need to connect a calibration tool or interface with process pH sensors.

  • Select Buffers: Two buffers bracketing your range (e.g., 4 and 7 for acidic; 7 and 10 for alkaline). These are suitable for sensors like Rika with a wide range (0-14pH).
  • Temperature Match: Ensure buffers are at process temp (25°C default). Use Rika's built-in compensation or manual input.
  • Rinse and Immerse in the First Buffer: Rinse the sensor. Immerse in pH 7 (or lowest). Stabilize 1-3 min (8-10s for Rika in flowing liquids).
  • Activate Calibration: Follow meter prompts. Confirm cleaning messages.
  • Rinse/Dry: Rinse with deionized water. Dab dry (no rubbing to prevent static).
  • Second Buffer: Repeat for pH 4/10. Meter adjusts slope/offset. Save data.
  • Verify: Re-check in buffers. Slope should be 92-102%. If there are warnings, such as a wrong buffer, reclean/retry.

 

Note: For multi-point, add 3+ buffers (e.g., 4, 7, 9, 10) for wide ranges or minimum uncertainty. Rika's high-precision models benefit from MPC to characterize nonlinearity.

Step 5: Post-Calibration and Verification

  • Document: Record date, buffers, slope/offset, temp.
  • Reinstall: Per application, like the process sensors from Rika submersible Type-B2.
  • Troubleshooting:
    • Noisy/Drift: Check wiring/ground loops. Clean again.
    • Nonlinearity: Expected in extremes and larger differences between the standard solutions. Use near-sample points.
    • Junction Mismatch/Sodium Error: Use double-junction references (Rika's options).

Step 6: Maintenance and Frequency for Long-Term Accuracy

  • Storage: In 3M KCl or saturated KCl. Avoid distilled water (dries electrode).
  • Frequency Adjustments: Inspect/clean every 3-6 months. Calibrate if deviation >±0.2 pH or post-extreme exposure.
  • Sensor Diagnostics: Modern meters (e.g., Rika's) monitor impedance/slope. Replace if impedance changes <92% or >102%.

Conclusion

Detecting pH in water provides insight into its chemical composition. It can detect changes that are otherwise invisible to the naked eye. Engineers and Scientists have developed various types of pH sensors, each applicable for specific uses. These include glass electrodes, combination electrodes, ISFETs, Metal Oxide Sensors, and submersible sensors. The most popular, cost-effective, and accurate design is glass electrodes. However, with time, the sensor detection capability can degrade. Therefore, a calibration frequency of 3 to 6 months is ideal for most industrial-grade sensors. Inspection, cleaning, calibration check, full calibration, and post calibration checks are key steps to perform during the calibration process. The process should restore the sensor's precision and accuracy.

 

Substandard sensors may require frequent calibration and may provide uneven responses over the exact buffer solutions. For high-accuracy and efficient recalibration, consider Rika water pH sensors. They provide a wide measurement range (0-14pH), exceptional resolution (0.01pH), temperature compensation (0-100°C), and power efficiency in their pH sensors (<0.15W). Moreover, their latest designs offer robust materials like Glass+316L or PC+ABS with IP68 ratings and 1MPa submersion. Visit Rika's website to explore all the options.

 

Frequently Asked Questions for pH Sensor

Q1:Why Calibrate a pH Sensor?

Every sensor will drift and lose its accuracy over time. The reason can be built on the sensing component. Therefore, the key is to clean and calibrate the sensor frequently. Typically, for industrial-grade pH sensors, 3-6months are enough. Regular calibration will help maintain sensor precision.

 

Q2:When to Calibrate Your pH Sensor

There are two approaches to calibration of pH sensors: preventive and corrective maintenance. Preventive maintenance involves periodic calibration and checks, which should occur every 3-6 months for industrial sensors and with every measurement for lab testing. Corrective maintenance is necessary when the pH sensor is exposed to extreme conditions, undergoes extended storage, yields inconsistent readings, or becomes contaminated.

 

Q3:What buffer solutions should I use?

Use certified reference materials (CRMs), span buffers with brackets of ≥2 pH units apart, match standard temperature 25°C, and use a neutral buffer. These will ensure accurate results, as the response may not be linear between different pH levels.

 

Can I calibrate in the field?

Yes, you can calibrate in the field. Portable pH meter, certified buffer solutions, distilled water, soap, detergent, tissues, 5% HCl/NaOH, and a stable surface. However, during the calibration process, it may be disconnected from the process controllers.

prev
The Differences between Rain Sensor and Rain Gauge
recommended for you
no data

LEAVE A MESSAGE

Just tell us your requirements, we can do more than you can imagine.
RIKA Sensor
Copyright © 2025 Hunan Rika Electronic Tech Co.,Ltd | Sitemap | Privacy Policy  
Customer service
detect