Turbidity Sensor for Water: How It Works, Uses & Selection Guide

RIKA SENSOR TECHNICAL GUIDE / WATER QUALITY MONITORING

Cloudy water is easy to see. But visual appearance alone cannot tell us how much suspended material is affecting the water. That is why turbidity measurement is an important part of water quality monitoring.

Turbidity is commonly monitored in wastewater treatment, surface water, aquaculture, industrial water and environmental monitoring applications.

For occasional testing, a portable turbidity meter may be sufficient. For continuous monitoring, an online turbidity sensor can measure water turbidity and send data to a controller, PLC or monitoring platform.

This guide explains what turbidity means, how optical turbidity sensors work, what NTU and FNU represent, where turbidity sensors are used and how to select the right sensor for your application.

QUICK ANSWER
A turbidity sensor for water measures an optical property associated with light scattering by suspended particles.

When choosing an online turbidity sensor, consider the expected NTU range, water type, measurement method, materials, installation conditions, fouling risk and communication requirements.

What Is Turbidity in Water?

Turbidity describes the cloudiness or haziness of water caused by suspended and colloidal particles that scatter light.

These particles may include:

Silt
Clay
Algae
Organic Matter
Industrial Solids
Comparison of clear water and turbid water showing differences in water clarity caused by suspended particles
Figure 1. Clear water and turbid water comparison. Suspended particles can affect water clarity by scattering light.

In general, water containing more light-scattering suspended material may appear more turbid.

However, turbidity is not a direct count of particles and does not identify their chemical composition.

Important: Turbidity Does Not Identify the Particles
A higher turbidity reading indicates a change in the water's optical condition. It does not, by itself, tell you whether the suspended material is clay, algae, corrosion products or another substance. Additional analysis may be needed to identify the cause.

Is Turbidity the Same as Total Suspended Solids (TSS)?

No. Although turbidity and TSS can be related, they represent different measurements.

OPTICAL MEASUREMENT
Turbidity

Measures an optical property related to light scattering by suspended particles in water. Common reporting units include NTU and FNU, depending on the method.

MASS-BASED MEASUREMENT
Total Suspended Solids

TSS is generally determined using a laboratory method that measures the mass concentration of retained suspended solids, commonly reported in mg/L.

Particle size, shape, color and optical properties influence light scattering. Therefore, two samples with similar TSS concentrations may produce different turbidity readings.

Turbidity should not automatically be converted into TSS without a validated relationship for the specific water being monitored.

How Does a Turbidity Sensor Work?

Most online water turbidity sensors use an optical measurement principle.

The RK500-07 uses a method based on detecting light scattered by particles in water at approximately 90° to the incident light.

OPTICAL TURBIDITY MEASUREMENT
Four Steps of Nephelometric Measurement
STEP 01
Light Emission
The sensor sends a controlled light beam into the water.
STEP 02
Particle Scattering
Suspended particles scatter some of the incident light.
STEP 03
Optical Detection
A detector measures scattered light at a defined angle.
STEP 04
Turbidity Output
Electronics process the signal and produce a calibrated turbidity reading.
Diagram explaining the 90-degree nephelometric turbidity sensor measurement principle using infrared light and scattered light detection
Figure 2. Illustration of the 90° nephelometric measurement principle used to detect light scattered by suspended particles.

This type of optical measurement is commonly called the nephelometric method.

Because it measures light scattering, the method can detect optical changes caused by suspended particles without requiring an operator to collect every sample manually.

Why Do Turbidity Sensors Use 860 nm Infrared Light?

The RK500-07 specification identifies an 860 nm infrared light source for its turbidity measurement system.

Near-infrared light around 860 nm is commonly associated with ISO 7027-type nephelometric measurement methods.

Compared with some visible-light techniques, this approach can reduce the effect of sample color on the optical reading.

LIGHT SOURCE
860 nm
Infrared optical measurement
DETECTION METHOD
90°
Nephelometric scattered-light detection

The selected wavelength and optical arrangement are important, but users should still confirm the measurement method required for their application, especially when the results will be used for regulatory or compliance reporting.

NTU vs FNU: What Do Turbidity Units Mean?

Two common units in turbidity measurement are NTU and FNU.

NTU
Nephelometric Turbidity Unit

Widely used for reporting turbidity measured by nephelometric instruments.

FNU
Formazin Nephelometric Unit

Commonly associated with ISO 7027-type infrared nephelometric methods.

Can NTU and FNU Be Used Interchangeably?
Not automatically. Results depend on the optical method, instrument design and applicable measurement standard. When comparing instruments or preparing compliance reports, confirm the required units and method.

For the RIKA SENSOR RK500-07, the current V5.0 technical parameter table specifies the selectable measurement ranges in NTU.

Online Turbidity Sensor vs Portable Turbidity Meter

Both instruments can measure turbidity, but they are intended for different monitoring needs.

PORTABLE / LABORATORY
Manual Spot Measurement
Suitable for occasional sampling, laboratory testing, site inspections and comparison measurements.
ONLINE / CONTINUOUS
Real-Time Trend Monitoring
Suitable for permanent or long-term installations, automatic data logging, alarms and industrial system integration.

One important advantage of online measurement is that operators can see how water turbidity changes over time, rather than relying only on isolated readings.

Where Are Turbidity Sensors Used?

Turbidity monitoring is used in a wide range of water-related applications. However, the purpose and required sensor design can differ significantly between them.

Illustrative online turbidity monitoring installation at an industrial wastewater treatment facility
Figure 3. Illustrative wastewater turbidity monitoring application. Actual installation should follow the selected sensor's installation requirements.

1. Wastewater Treatment

Wastewater may contain suspended solids, biological material, flocs and other particles.

Turbidity measurement can help operators observe changes in water clarity during suitable treatment stages, including clarification, filtration or treated-water monitoring.

However, turbidity is not a direct substitute for TSS or other wastewater parameters. The measurement should be selected according to the monitoring objective.

2. Industrial Wastewater

Industrial wastewater conditions differ between facilities. Some processes contain suspended minerals, while others may involve organic contamination, chemicals or corrosive substances.

For these applications, the sensor must be selected according to the expected turbidity range, wetted-material compatibility, fouling conditions, temperature and pressure.

3. River and Surface Water Monitoring

Rivers, lakes, reservoirs and other surface waters may experience turbidity changes associated with rainfall, runoff, erosion, sediment disturbance and biological activity.

Continuous measurement can help environmental monitoring teams observe these changes and develop longer-term records.

For unattended stations, waterproofing, power supply, output interface and maintenance access are also important.

4. Aquaculture

In aquaculture, suspended particles may come from sediment, uneaten feed, organic material or other biological activity.

Turbidity can be used as one part of a wider water-quality monitoring program.

Depending on the system, parameters such as dissolved oxygen, temperature, pH and conductivity may also be relevant.

5. Mariculture and Corrosive Water

Marine environments create different material requirements from freshwater.

The RK500-07 series includes a titanium-alloy Type B configuration intended for applications such as mariculture, strongly corrosive sewage and other complex conditions.

This highlights why material compatibility is an important selection factor alongside turbidity range.

6. Pipeline Turbidity Monitoring

Some industrial systems require turbidity measurement directly in a pipeline rather than in an open tank or basin.

The RK500-07 Type C is designed for pipeline installation using 316L and UPVC materials.

The current product drawing indicates a connection designed for UPVC with an outer diameter of 32 mm.

In pipeline applications, bubbles, reflective surfaces, flow conditions and the position of the optical measurement area should be considered.

7. Clean-in-Place (CIP) Applications

The RK500-07 product specification also lists Clean-in-Place applications.

However, actual compatibility must be confirmed against the cleaning chemical, concentration, temperature, pressure and cleaning procedure.

Application Note
Listing CIP as an application does not mean that every sensor configuration is suitable for all cleaning chemicals or all CIP conditions. Confirm compatibility before selection.

How to Choose a Turbidity Sensor for Water

Choosing a turbidity sensor is not simply about finding an instrument that can measure NTU.

A sensor suitable for relatively clear water may not be the best choice for highly turbid wastewater, seawater or a fouling-prone process.

SENSOR SELECTION CHECKLIST
Six Key Factors Before Selection
01
Measurement Range
What is the expected minimum, normal and maximum turbidity?
02
Water Type
Freshwater, wastewater, seawater, surface water or industrial liquid?
03
Material Compatibility
Are salt, corrosive substances or cleaning chemicals present?
04
Fouling Conditions
Will biofilm, sediment or sludge accumulate on the optical window?
05
Installation Method
Is the sensor installed in a tank, river, channel or pipeline?
06
Communication Output
Does the control system require 4–20 mA, RS485 or both?

What Turbidity Range Should You Choose?

Range selection is one of the most important decisions.

Relatively clear water may require good performance at low turbidity, while wastewater and sediment-rich water may require a wider range.

The RK500-07 offers the following selectable ranges:

0–10 NTU
0–100 NTU
0–1000 NTU
0–4000 NTU

A wider range is not automatically better. The selected range should match the actual process conditions and required measurement performance.

Should You Choose a Self-Cleaning Turbidity Sensor?

Optical measurement requires a clean optical window.

In wastewater, aquaculture or other fouling-prone applications, sediment, biological growth or deposits may accumulate on the measurement surface.

The RK500-07 Type D includes an integrated cleaning brush designed to help reduce fouling on the optical measurement area.

Self-Cleaning Does Not Mean Maintenance-Free
A cleaning brush can reduce manual cleaning requirements in suitable applications. However, the optical surface still requires inspection and cleaning when necessary.

How to Install an Online Turbidity Sensor

Good installation is essential for stable optical turbidity measurement.

The installation should provide representative water conditions while minimizing interference from bubbles, deposits, reflective surfaces and other disturbances.

Immersion Installation
The sensor is placed directly into the liquid, such as in a tank, basin, river or open channel.
Pipeline Installation
A suitable sensor configuration is installed in a pipe or process connection for continuous measurement.
Bypass / Flow Cell
Where the system design permits, a bypass arrangement may improve access for inspection and maintenance.

Common Installation Factors That Affect Readings

Air Bubbles
Bubbles may scatter light and interfere with the optical reading. Avoid locations with persistent air accumulation where possible.
Optical Fouling
Deposits, biofilm or sediment on the optical window can affect light transmission and scattering.
Reflective Surfaces
Nearby reflective pipe walls may influence the optical measurement and require appropriate clearance or orientation.
Particle Characteristics
Particle size, shape and optical properties influence scattering and can affect the relationship between turbidity and solids concentration.

Special Considerations for Reflective Pipelines

Stainless-steel and other reflective pipe surfaces may influence optical readings when the measurement area is too close to the wall.

The RK500-07 V5.0 installation guidance recommends maintaining at least 58 mm clearance from a reflective surface in the illustrated arrangement.

For reflective pipelines, the document also recommends aligning the line between the two small optical openings with the pipeline flow direction to reduce reflection effects.

Installation Reminder
Installation geometry should be checked against the selected sensor configuration and the current manufacturer's installation drawing. Do not automatically apply one clearance or mounting arrangement to every turbidity sensor design.

How Do You Maintain a Turbidity Sensor?

Because a turbidity sensor uses an optical measurement system, maintaining a clean optical window is important.

The RK500-07 specification recommends keeping the optical surface clean and wiping it gently when necessary.

Maintenance Checklist
01. Inspect the optical window for sediment, biofilm and other deposits.
02. Clean the measurement surface gently using an appropriate method.
03. Avoid scraping the optical surface with hard objects.
04. Avoid strong vibration and mechanical damage.
05. Check sensor readings and perform calibration or verification according to the applicable procedure and operating requirements.

The actual maintenance frequency depends on water conditions. Heavily contaminated wastewater or biofouling-prone environments may require more frequent inspection than relatively clear water.

RIKA SENSOR RK500-07 Turbidity Sensor: Four Configurations

The RK500-07 series provides four configurations for different water environments, installation conditions and maintenance requirements.

RIKA SENSOR RK500-07 turbidity sensor Type A Type B Type C and Type D configurations for freshwater corrosive water pipeline installation and self-cleaning applications
Figure 4. RK500-07 turbidity sensor configurations: Type A for general freshwater, Type B for corrosive water, Type C for pipeline installation and Type D with an integrated cleaning brush.
Type Specified Application Main Material Key Feature
Type A General freshwater environments 316L stainless steel General-purpose water monitoring
Type B Mariculture, strongly corrosive sewage and complex conditions Titanium alloy Corrosion-resistant construction
Type C Pipeline installation 316L + UPVC Pipeline-oriented installation structure
Type D Applications requiring an integrated cleaning brush 316L stainless steel Self-cleaning / reduced-maintenance design

RK500-07 Technical Specifications

The following specifications are based on the RK500-07 Turbidity Sensor V5.0 document. Values that differ by configuration are identified separately.

Parameter Specification
Measurement Principle Optical, 90° scattered-light measurement
Light Source 860 nm infrared
Measurement Ranges 0–10 / 0–100 / 0–1000 / 0–4000 NTU
Accuracy Types A/B/C: ±1% FS; Type D: ±2% FS
Resolution Types A/B/C: 0.01 NTU at ≤10 NTU; 0.1 NTU above 10 NTU. Type D: 0.1 NTU
Response Time 1 s
Power Supply 7–30 VDC
Output 4–20 mA and RS485 simultaneously
Ingress Protection IP68
Operating Temperature Types A/B/C: 0 to +60°C; Type D: -5 to +60°C
Specified Pressure Resistance 0.8 MPa nominal rating in the specification; confirm permitted working pressure for the selected version before installation
Default Cable Length 5 m
Optical Window Sapphire measurement window
Self-Cleaning Type D includes a cleaning brush
Specification Reminder
Measurement range, mechanical configuration, accuracy, operating temperature and other relevant conditions should be checked for the actual selected type. The information above does not mean that every RK500-07 configuration has identical performance or installation requirements.

Can One Turbidity Sensor Measure Both Clean Water and Wastewater?

Not necessarily.

Low-turbidity water and highly turbid wastewater can create very different measurement requirements.

Relatively Clear Water
May require greater attention to low-range resolution, optical stability and the measurement method.
Highly Turbid Water
May require a wider measurement range, suitable materials, fouling resistance and more frequent maintenance.

The correct sensor should match the expected turbidity range, water characteristics, installation and required performance.

What Should You Tell a Turbidity Sensor Supplier Before Buying?

Preparing the following project information helps make sensor selection faster and more accurate.

01. Water Type Freshwater, wastewater, seawater or another medium?
02. Turbidity Range What are the normal and maximum NTU values?
03. Installation Tank, river, open channel or pipeline?
04. Temperature What is the operating temperature range?
05. Pressure Will the sensor operate inside a pressurized system?
06. Corrosion Are salt or aggressive chemicals present?
07. Fouling Is biofilm, sludge or sediment accumulation expected?
08. Output Signal 4–20 mA, RS485 or both?

Frequently Asked Questions

What does a turbidity sensor measure?

It measures an optical property associated with light scattering by suspended particles in water. It does not directly identify the particles or their chemical composition.

What does NTU mean?

NTU stands for Nephelometric Turbidity Unit. It is commonly used to report turbidity measurements obtained using nephelometric methods.

Is higher NTU always a sign of unsafe water?

No. Higher turbidity generally indicates more light scattering, but turbidity alone does not determine water safety or identify the type of contamination.

Is turbidity the same as TSS?

No. Turbidity is an optical measurement, while TSS is generally a mass-based measurement. A site-specific relationship is needed before using turbidity to estimate TSS.

Can a turbidity sensor continuously monitor water?

Yes. Online turbidity sensors can be installed for continuous monitoring when the selected configuration and installation conditions are suitable.

Can a turbidity sensor be installed in a pipeline?

Yes, provided the sensor is designed for pipeline installation. The RK500-07 Type C has a pipeline-oriented configuration using 316L and UPVC.

Does a self-cleaning turbidity sensor need maintenance?

Yes. A cleaning brush can reduce fouling and manual cleaning requirements, but inspection and cleaning may still be necessary.

What range should I choose for an online turbidity sensor?

Choose a range that covers the expected water conditions while meeting the required measurement performance. RK500-07 offers selectable 0–10, 0–100, 0–1000 and 0–4000 NTU ranges.

Can an online turbidity sensor connect to a PLC?

Yes, if the interfaces are compatible. RK500-07 provides 4–20 mA and RS485 outputs according to its current technical specification.

Can turbidity measurement identify corrosion in a pipeline?

Not directly. Suspended corrosion products may affect turbidity, but a turbidity reading alone cannot prove corrosion, identify particle composition or determine a corrosion rate.

Final Thoughts: The Best Turbidity Sensor Depends on the Water

When choosing a turbidity sensor, it is easy to compare only measurement range, accuracy and price.

But reliable continuous turbidity measurement depends on more than those specifications.

KEY TAKEAWAY
Select a turbidity sensor based on the expected water conditions, measurement range, optical method, material compatibility, installation environment, fouling risk, maintenance needs and output requirements.

A sensor that works well in freshwater may not be ideal for seawater, heavily contaminated wastewater or a reflective pipeline.

For online monitoring, the goal is not simply to display an NTU value. It is to obtain a stable, maintainable measurement that is representative of the water being monitored.

Online Turbidity Monitoring with RIKA SENSOR

RIKA SENSOR provides online turbidity measurement solutions for environmental monitoring, aquaculture, wastewater treatment and industrial water applications.

The RK500-07 series includes configurations for general freshwater, corrosive water, pipeline installation and applications where an integrated cleaning brush can help reduce fouling.

With multiple turbidity ranges, IP68 protection and 4–20 mA plus RS485 outputs, the sensors can be integrated into compatible water-quality monitoring and industrial control systems.

WATER QUALITY MONITORING / RIKA SENSOR

Need Help Selecting a Turbidity Sensor for Water?

Tell us your water type, expected NTU range, operating temperature and pressure, installation method, fouling conditions and communication requirements. RIKA SENSOR can help evaluate the appropriate turbidity sensor configuration for your project.

prev
pH Sensor for Water: How It Works, Applications and Selection Guide
Dissolved Oxygen in Water: Optical DO Sensors Explained
next
recommended for you

LEAVE A MESSAGE

Just tell us your requirements, we can do more than you can imagine.
RIKA Sensor
Copyright © 2026 Hunan Rika Electronic Tech Co.,Ltd | Sitemap | Privacy Policy  
Customer service